Home About Maccor Products Lab Equipment Series 4000 Model 4200 Model 4300 Model 4304 Series 4600 Primary Cell Testers Series 3600 Series 3650 HEV, PHEV, EV Tester Formation Electrochemical Software Options Accessories Temperature Chamber Why Choose Maccor? Contact Us Worldwide |
SERIES 8000 HIGH POWER AUTOMATED TEST SYSTEMThe Series 8000 is a range of fully automated, computerized battery test systems that provide the highest level of specifications and features available in a Maccor test system. The Series 8000 provides exceptional accuracy (99.95% of Full Scale), resolution (16-bit, 1 part in 65,536) and speed (1 mS data aquisition, available) for testing Hybrid Electric Vehicle (HEV) batteries, Super Capacitor, and other high power energy storage devices. With channel ratings as high as 700 Volts and 1000 Amps, the Series 8000 is the perfect instrument in your toolbox for high power applications. Using Maccor's Standard Full Feature Battery Test Software, the Series 8000 is the most powerful test system you will own. Perform virtually any type of test (FUDS, SFUDS, DST, etc) on any type of energy storage device (batteries, battery packs, super-capacitors, fuel cells, etc.). The Series 8000 comes standard with the high level accuracy, resolution, and dependability satisfied Maccor customers expect. Supplied as a complete turnkey system, the Series 8000 consists of a test cabinet, PC computer and test software, with options of a UPS and Printer. The test cabinet and PC computer are connected via a high-speed 10 Base T communications network. As soon as AC power is connected to the test system, it is ready for immediate use. To provide maximum flexibility and speed of operation, single board computer controllers within the test cabinet provide control of the tests and collect the initial data. Each controller supports from one to eight test channels, dependent on the application. In addition, the test cabinet contains the individually controlled electronic loads, and power supplies for charging purposes. Each test channel is operated independently. This allows different tests to be performed on different test channels simultaneously, with tests being started independently, or in a block. Once started, tests operate completely automatically until the appropriate test end condition is reached.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Configuration | ||
|
Charge Current |
increments of 145 Amps |
|
Discharge Current |
increments of 100 Amps |
|
Voltage |
increments of 50 volts, maximum of 700 volts |
|
Power |
increments of 50 kW |
Accuracy | ||
|
Current |
0.05% of Full Scale |
|
Voltage |
0.05% of Full Scale |
Resolution | ||
|
Current |
16-bit (1 part in 65,536) |
|
Voltage |
16-bit (1 part in 65,536) |
Time | ||
|
Minimum Step Time |
10 mS |
|
Control, Measurement, Adjustment |
10 mS |
|
Data Collection |
10 mS Standard, 1 mS Option |
Modes of Operation | ||
|
Constant |
Current, Power, Voltage, and Resistance |
|
Rest | |
|
Import Test Step: Where an ASCII text file containing a table (list) of constant current or power values at fixed or specified intervals is executed as a test step. The values in the table can be automatically scaled for W/kg or C rate. | |
1kHz High-Speed Data Acquisition Burst Mode | ||
Auxiliary Voltage Inputs |
±5V, ±10V, ±20V, etc. | |
Temperature Monitoring |
K, J, T, thermistor | |
Control of External Load and/or Charger | ||
Environmental Chamber Control | ||
All specifications subject to change, please contact Maccor for latest specification.
© 2009 Maccor, Inc. All Rights Reserved
Maccor, Inc.
4322 South 49th West Avenue
Tulsa, OK 74107 - USA
Tel: 918 446 1874
Fax: 918 445 1496
e-mail: sales@maccor.com