MACCOR - Primary Cell Testers

MACCOR - Battery and Cell Test Equipment Worldwide
 

+ Primary Cell Testers

The multifunction Series 3600 primary cell test system has been engineered to meet the growing demands of existing and upcoming ANSI and IEC test standards. Each test position: Operates independently of the other positions; Can be user programmed to run a wide range of multi-step tests, including pulse tests; Can be user programmed to operate in steps of fixed current, fixed power, fixed voltage, fixed resistance; Has a dynamic range of over 25,000 to 1 to meet ANSI and IEC standards of current|power|resistance.

The Series 3600 is supplied as a complete turnkey system with PC computer and software. Each system is available from a desktop model to large test and quality control models with thousands of test positions. Desktop models are usually available from stock.

Powered by Maccor’s standard test system software and utilizing similar technology as found in Maccor’s high-precision Series 4000 Automated Test System, the Series 3600’s are ideal for today’s primary cell testing standards. The software can be configured to automatically print a standard primary cell end of test report. Custom data handling interfaces or user interfaces are available as options.

The Series 3600 has a Phoenix connector interface with horizontal cell configuration providing a dense and compact configuration. Each cabinet (standard 19” rack) can handle up to 192 test positions. The Series 3600 is also available in a desktop cabinet with 32-test positions.

The Series 3650 has a Powerpole® interface with vertical cell configuration. The 3650 has a slightly larger footprint than the 3600, each cabinet can be configured with as many as 256 test positions.

Specifications

Number of Test Channels: 32 to 256 per cabinet
Voltage Measurement Range:0 V to +5 V
Voltage Accuracy:±1 mV
Minimum Discharge Voltage:0.2 volts or 0.35 × Current, whichever is greater
Voltage Resolution:16 bit
Current
Range 1:1 mA Full Scale ±0.2 µA
Range 2:15 mA Full Scale ±3.0 µA
Range 3:200 mA Full Scale ±40.0 µA
Range 4:2500 mA Full Scale ±500 µA
Maximum Charge Current:none
Current Resolution:16 bit
Time Resolution:50 mS
Minimum Step Time:500 mS
Data Recording Rate:200 data points per second per system standard
Operating Modes:Constant Current, Constant Voltage, Constant Power, Constant Resistance
Number of Steps per Test:18 standard plus subroutines
 
sales@maccor.com    +1 918 446 1874
 
© 2018 Maccor, Inc. All Rights Reserved